Selected:

DeFelsko PosiTector® 6000 Coating Thickness Gage

Product Information for:

DeFelsko PosiTector® 6000 Coating Thickness Gage

PosiTector 6000 Features

Magnetic film gages are used to nondestructively measure the thickness of a nonmagnetic coating on ferrous substrates. Most coatings on steel and iron are measured this way. Magnetic gages use one of two principles of operation: magnetic pull-off or magnetic/electromagnetic induction.

Rugged, fully electronic coating thickness gauges use magnetic and eddy current principles to measure coating thickness on both ferrous and non-ferrous metals, accurately and quickly.

Eddy current techniques are used to nondestructively measure the thickness of nonconductive coatings on nonferrous metal substrates. A coil of fine wire conducting a high-frequency alternating current (above 1 MHz) is used to set up an alternating magnetic field at the surface of the instrument’s probe. When the probe is brought near a conductive surface, the alternating magnetic field will set up eddy currents on the surface. The substrate characteristics and the distance of the probe from the substrate (the coating thickness) affect the magnitude of the eddy currents. The eddy currents create their own opposing electromagnetic field that can be sensed by the exciting coil or by a second, adjacent coil.
  1. Simple
    • Ready to measure-no calibration adjustment required for most applications
    • One-Handed menu navigation
    • Bi-color indicator light-ideal in a noisy environment
    • RESET feature instantly restores gauge to factory settings
  2. Durable
    • Solvent, acid, oil, water and dust resistant-meets or exceeds IP5X ingress protection
    • Wear-resistant probe tip
    • Thick, impact and scratch resistant Lexan® display suitable for harsh environments
    • Protective rubber holster with quick-release belt clip and probe storage
    • Two (2) year warranty on gage body and probe.
  3. Accurate
    • Certificate of Calibration showing traceability to NIST included
    • Built-in temperature compensation ensures measurement accuracy
    • Hi-RES mode increases gages displayed resolution for use on applications that require greater precision
    • Conforms to national and international standards including ISO and ASTM
  4. Versatile
    • Multiple calibration adjustment options including 1 point, 2 point, known thickness and average zero
    • Calibration Lock
    • SSPC PA2 feature set determines if film thickness over a large area conforms to user-specified min/max levels (memory (3) gauges only)
    • PSPC 90/10 feature determines if a coating system complies with an IMO performance standard for protective coatings (memory (3) gages only)
    • Selectable display languages
    • Flip Display for right-side-up viewing
    • Backlit display for dim or dark environments
    • Mils/Microns/mm switchable
    • Extended cables available (up to 250 ft/75 m) for remote measuring
    • All Regular Separate Probes suitable for underwater use
    • All PosiTector 6000 Gauges are available in your choice of Basic, Standard and Memory configurations
    • Menu-driven user interface with multilingual support including: English, Spanish, French, German, Portuguese, Chinese, Japanese, Norwegian, Russian, Czech and Italian – Gauge Screen Captures

Three Steps to Selecting a PosiTector 6000

  1. Select Substrate
    • F – for ferrous metals (steel and cast iron)
    • N – for non-ferrous metals (aluminum, copper, etc.)
    • FN – for all metal substrates — Gage automatically recognizes the substrate and takes a measurement
  2. Select Measurement Range and Probe Style
    • Removable built-in probe can be detached and replaced with any one of a wide variety of probes. Available ranges and probe styles are shown in the ordering guide below.
  3. Select Body Model
    • Basic Model
    • Standard Model
    • Memory Model
    • Features of each model are listed in this features comparison chart.
Positector 6000 Feature Comparison Chart
FeatureBasic ModelStandard ModelMemory Model
No Calibration required for most applications
One handed menu navigation
Average Zero Feature; ideal for zeroing on rough substrates
Mils/Microns/MM switchable
Adjustable to a known thickness
Thick impact resistant Lexan® display
Flip Display enables right side up viewing whether the
gage is in your hand or on a work table
Backlit display for dim or dark environments
High Resolution mode increases the displayed resolution
for use on applications that require greater precision
Hi-Lo limits audibly and visibly alerts when measurements
exceed your specified limits
Instantly calculate average, standard deviation, min/max thickness
and number of readings while measuring
Reading Storage Capacity250 in
one group
10,000
in up to
1000 groups
SSPC PA2 feature set determines if film thickness over
a large area conforms to user-specified min/max levels
PSPC 90/10 feature determines if a coating system complies
with an IMO performance standard for protective coatings
BlueTooth logo Wireless Technology for immediate data
transfer to a PC or Bluetooth® compatible device
Built in clock to date and time stamp each stored measurement
Multiple stored calibration adjustments for measuring
on a variety of substrate conditions
PosiSoft downloading and charting software
(includes USB cable)
optional
Built-in IR port for printing to an optional portable printer
Protective rubber holster with belt clip
Nylon carry case with shoulder strap
Certificate of Calibration showing traceability to NIST
Two (2) Year Warranty
All gages come complete with precision plastic shims, protective rubber holster with belt clip, 3 AAA batteries, instructions, wrist strap (built-in style only), nylon carrying case with shoulder strap, Certificate of Calibration traceable to NIST, two (2) year warranty.

 

PROBE DETAILS

 

Positector 6000 probes

Removable Built-in Probes
Probe ModelFNFNFT
Positector 6000 removeable built-in probeRange0 – 60 mils0 – 250 mils
0 – 1500 μm0 – 6 mm
Accuracy*± (0.05 mils + 1%) 0 – 2 mils
± (0.1 mils + 1%) > 2 mils
± (0.5 mils + 1%) 0 – 100 mils
± (0.5 mil + 3%) > 100 mils
± (1 μm + 1%) 0 – 50 μm
± (2 μm + 1%) > 50 μm
± (0.01 mm + 1%) 0 – 2.5 mm
± (0.01 mm + 3%) > 2.5 mm
*Accuracies are stated as a fixed value plus a percentage of the gage’s actual reading.
1 mil = 0.001 inch (1 X 10-3 inch)       1 mm = 0.001 m. (1 X 10-3 m)
1 μm = 0.000001 m. (1 X 10-6 m)
Separate Regular Probes
RegularProbe ModelFSNSFNSNASFTS
Positector 6000 removeable built-in probeRange0 – 60 mils0 – 25 mils0 – 250 mils
0 – 1500 μm0 – 625 μm0 – 6 mm
Accuracy*± (0.05 mils + 1%) 0 – 2 mils
± (0.1 mils + 1%) > 2 mils
± (0.02 mils + 1%) 0 – 4 mils
± (0.1 mils + 3%) > 4 mils
± (0.5 mils + 1%) 0 – 100 mils
± (0.5 mil + 3%) > 100 mils
± (1 μm + 1%) 0 – 50 μm
± (2 μm + 1%) > 50 μm
± (5 μm + 1%) 0 – 100 μm
± (2 μm + 3%) > 100 μm
± (0.01 mm + 1%) 0 – 2.5 mm
± (0.01 mm + 3%) > 2.5 mm
*Accuracies are stated as a fixed value plus a percentage of the gage’s actual reading.
1 mil = 0.001 inch (1 X 10-3 inch)       1 mm = 0.001 m. (1 X 10-3 m)
1 μm = 0.000001 m. (1 X 10-6 m)
Separate Regular Probes
90° RegularProbe ModelFRSNRSFNRS
Positector 6000 removeable built-in probeRange0 – 60 mils
0 – 1500 μm
Accuracy*± (0.05 mils + 1%) 0 – 2 mils
± (0.1 mils + 1%) > 2 mils
± (1 μm + 1%) 0 – 50 μm
± (2 μm + 1%) > 50 μm
*Accuracies are stated as a fixed value plus a percentage of the gage’s actual reading.
1 mil = 0.001 inch (1 X 10-3 inch)       1 μm = 0.000001 m. (1 X 10-6 m)
Separate Microprobes
0° MicroprobeProbe ModelF0SN0S
Positector 6000 removeable built-in probePositector 6000 removeable built-in probeRange0 – 45 mils0 – 25 mils
0 – 1150 μm0 – 625 μm
Accuracy*± (0.02 mils + 1%) 0 – 4 mils
± (0.1 mils + 3%) > 4 mils
± (5 μm + 1%) 0 – 100 μm
± (2 μm + 3%) > 100 μm
*Accuracies are stated as a fixed value plus a percentage of the gage’s actual reading.
1 mil = 0.001 inch (1 X 10-3 inch)       1 μm = 0.000001 m. (1 X 10-6 m)
Separate Microprobes
45° MicroprobeProbe ModelF45SN45S
Positector 6000 removeable built-in probeRange0 – 45 mils0 – 25 mils
0 – 1150 μm0 – 625 μm
Accuracy*± (0.02 mils + 1%) 0 – 4 mils
± (0.1 mils + 3%) > 4 mils
± (5 μm + 1%) 0 – 100 μm
± (2 μm + 3%) > 100 μm
*Accuracies are stated as a fixed value plus a percentage of the gage’s actual reading.
1 mil = 0.001 inch (1 X 10-3 inch)       1 μm = 0.000001 m. (1 X 10-6 m)
Separate Microprobes
90° MicroprobeProbe ModelF90SN90S
Positector 6000 removeable built-in probeRange0 – 45 mils0 – 25 mils
0 – 1150 μm0 – 625 μm
Accuracy*± (0.02 mils + 1%) 0 – 4 mils
± (0.1 mils + 3%) > 4 mils
± (5 μm + 1%) 0 – 100 μm
± (2 μm + 3%) > 100 μm
*Accuracies are stated as a fixed value plus a percentage of the gage’s actual reading.
1 mil = 0.001 inch (1 X 10-3 inch)       1 μm = 0.000001 m. (1 X 10-6 m)
Separate High-Range Probes
High-RangeProbe ModelFNTSFKSNKS
Positector 6000 removeable built-in probeRange0 – 250 mils0 – 500 mils
0 – 6 mm0 – 13 mm
Accuracy*± (0.05 mils + 1%) 0 – 100 mils
± (0.5 mils + 3%) > 100 mils
± (1 mils + 3%) 0 – 500 mils
± (0.02 mm + 3%) 0 – 13 mm
± (0.01 mm + 1%) 0 – 2.5 mm
± (0.01 mm + 3%) > 2.5 mm
*Accuracies are stated as a fixed value plus a percentage of the gage’s actual reading.
1 mil = 0.001 inch (1 X 10-3 inch)       1 μm = 0.000001 m. (1 X 10-6 m)

 

ORDERING INFORMATION

PosiTector® 6000 Series Ordering Chart
ORDERING
CHART
Build-in and
Regular separate
probes
90° Regular
probe for
for tight spots
Ideal for
anodized
aluminum
Microprobes
for small parts or
hard to reach spots
Built-in and separate
probes for thick coatings,
epoxy, rubber, etc.
Positector 6000Positector 6000Positector 6000Positector 6000Positector 6000Positector 6000Positector 6000Positector 6000Positector 6000Positector 6000
F
E
R
R
O
U
S
Gage with Probe

Basic

Standard

Memory

Probe Only

Coatings on FERROUS metals (Steel & Cast Iron)
F1FS1FRS1F0S1F45S1F90S1FT1FTS1FKS1
F2FS2FRS2F0S2F45S2F90S2FT2FTS2FKS2
F3FS3FRS3F0S3F45S3F90S3FT3FTS3FKS3
FFSFRSF0SF45SF90SFTFTSFKS
    F
E
N R
O R
N O
U
S
Gage with Probe

Basic

Standard

Memory

Probe Only

Coatings on NON-FERROUS metals (Aluminum, Copper, etc.)
N1NS1NRS1NAS1N0S1N45S1N90S1NKS1
N2NS2NRS2NAS2N0S2N45S2N90S2NKS2
N3NS3NRS3NAS3N0S3N45S3N90S3NKS3
NNSNRSNASN0SN45SN90SNKS
C
O
M
B
O
Gage with Probe

Basic

Standard

Memory

Probe Only

Combinations ALL METALS metals (Ferrous & Non-Ferrous)
FN1FNS1FNRS1FNTS1
FN2FNS2FNRS2FNTS2
FN3FNS3FNRS3FNTS3
FNSFNSFNRSFNTS
Measuring
Range
0 – 60 mils

0 – 1500 μm

Ferrous:
0 – 45 mils (0 – 1500 μm)Non-Ferrous:
0 -25 mils (0 – 626 μm)
0 – 250 mils

0 – 6 mm

0 – 500 mils

0 – 13 mm

Accuracy*± (0.05 mils + 1%) 0 – 2 mils
± (0.1 mils + 1%) > 2 mils± (1 ±m + 1%) 0 – 50 μm
± (2 μm + 1%) > 50 μm
± (0.02 mils + 1%) 0 – 4 mils
± (0.1 mils + 3%) > 4 mils±(0.5 μm + 1%) 0 – 100 μm
± (2 μm + 3%) > 100 μm
±( 0.5 mils + 1%)
0 – 100 mils
±( 0.5 mils + 3%)
> 100 mils± (0.01 mm + 1%) 0 – 2.5 mm
±( 0.01 mm + 3%) > 2.5 mm
± (1 mil + 3%)
0 – 500 mils± (0.02 mm + 3%)
0 -13 mm
Matching DeFelsco
Calibration Standards
S1

A1

S2

A2

P1P5
*Accuracies are stated as a fixed value plus a percentage of the gage’s actual reading.

 

What’s Included
Positector 6000_incudes

All gages come complete with precision plastic shims, protective rubber holster with belt clip, 3 AAA batteries, instructions, wrist strap (built-in style only), nylon carrying case with shoulder strap, Certificate of Calibration traceable to NIST, two (2) year warranty.

Memory (3) gages also include PosiSoft Windows-based Analysis Software, USB cable.

Size: 146 x 64 x 31 mm (5.75″ x 2.5″ x 1.2″)
Weight: 165 G (5.8 oz.) without batteries

Conforms to national and international standards ISO 2178/2360/2808, PrEN ISO 19840, ASTM B499/D1186/D1400/D7091/E376/ G12, BS3900-C5, SSPC-PA2 and others

Additional information

SKU DF-PT6000 Categories , , Tags ,
Share on facebook
Share on linkedin
Close Menu
×